Electrochemical Nitrogen Dioxide Sensor NO2-MD-600
Design Features
- Quick Response
- Excellent Selectivity
- Linearity
- Stability
- High Reliability
- Perfect Leak-proof Structure
Specifications [Sensitivity Characteristics]
Detection Gas | Nitrogen Dioxide |
---|---|
Detection Range | 0 ~ 30 ppm |
Maximum Overload | 150 ppm |
Output Signal | 600± 150 nA/ppm |
Repeatability | ± 2 % |
Resolution | 0.1 ppm |
Typical Baseline Range (pure air) | ±0.2 ppm |
Typical Response Time (t90) | < 25 seconds |
Baseline Shift (- 20 ~ 40℃ ) | < 0.2 ppm |
Long Term Output Drift | < 2 % / month |
Expected Life Time | > 3 years |
Performance data conditions: 20℃, 50%RH and 1013mBar.
Operating Conditions
Operating Temperature | -20~50℃ |
---|---|
Operating Humidity | 15~90%RH |
Operating Pressure Range | Atmospheric± 10 % |
Recommended Load Resistor | 33Ω |
Bias Voltage | Not required |
Position Sensitivity | None |
Recommended Storage Temp. | 0~20℃ |
Storage Life | 6 months |