Electrochemical Sulfur Dioxide Sensor SO2-MD-200
Design Features
- Measurement for High Concentration
- Excellent Selectivity
- Linearity
- Stability
- High Reliability
- Perfect Leak-proof Structure
Specifications [Sensitivity Characteristics]
Detection Gas | Sulfur Dioxide |
---|---|
Detection Range | 0 ~ 100 ppm |
Maximum Overload | 500 ppm |
Output Signal | 200± 50 nA/ppm |
Repeatability | ± 2 % |
Resolution | 0.2 ppm |
Typical Baseline Range (pure air) | -0.5 ppm to +1.5 ppm |
Typical Response Time (t90) | < 30 seconds |
Baseline Shift (- 20 ~ 40℃ ) | < 0.5 ppm |
Long Term Output Drift | < 2 % / month |
Expected Life Time | > 2 years |
Performance data conditions: 20℃, 50%RH and 1013mBar.
Operating Conditions
Operating Temperature | -30~50℃ |
---|---|
Operating Humidity | 15~90%RH |
Operating Pressure Range | Atmospheric± 10 % |
Recommended Load Resistor | 10Ω |
Bias Voltage | 0mV |
Position Sensitivity | None |
Recommended Storage Temp. | 0~20℃ |
Storage Life | 6 months |